Also available are consumables for
– AVL 9180
– AVL 9181
– STARLYTE III
– STARLYTE V ELECTROLYTE ANALYZERS
 These consumables include
– Electrodes
– Electrode conditioner
– Cleaning solution
Na+: Electrode
K+: Electrode
Cl-: Electrode
Ca++:Electrode
Li+: Electrode

Specification

Sample
Whole Blood, Serum, Plasma, Urine, or Dialysate

Sample Size
95 μl

Detection Range
Blood Na+: 40-200 mmol/ L
Dialysate K+: 1.7-15 mmol/L
Cl-: 50-200 mmol/ L
Ca++: 0.3-5.0 mmol/L
Li+: 0.2-5.5 mmol/L
Urine Na+: 3-300 mmol/L
K+: 5-120* mmol/L
Cl-: 15-300 mmol/L
*60-120 requires additional dilution

Display Resolution
Na+: 1 mmol/L or 0.1 mmol/L
K+: 0.1 mmol/L or 0.01 mmol/ L
Cl-: 1 mmol/L or 0.1 mmol/L
Ca++: 0.01 mmol/L or 0.001 mmol/L
Li+: 0.01 mmol/L or 0.001 mmol/L

Analysis Time
60 sec without printout (60 per hou r)
75 sec with printout (48 per hour)

Data Storage
1000 Patient Results
500 Level 1 QC Results
500 Level 2 QC Results
500 Level 3 QC Results

Calibration
2 point every 4 hours
1 Point after every Measurement

Reproducibility
Within Run (n=30 )
Blood, Serum, Plasma, Dialysate
Na+: CV  1% (120-160 mmol/ L)
K+: CV  2% (2.5-6 mmol/L )
Cl-: CV  2% (85-130 mmol/ L)
Ca++: SD  0.02 mmol/L (0.8 – 1 .5 mmol/L)
Li+: SD  0.02 mmol/L (0.4 – 1 .3 mmol/L)
(Lithium not typically measured in dialysate samples)
Urine
Na+: CV  5% (100-250 mmol/ L)
K+: CV  5% (10-60 mmol/L )
Cl-: CV  5% (100-250 mmol/ L)
(Calcium and Lithium are not typically measured in urine samples)

Veterinary Options
Feline, Canine, Bovine, Equine, Swine, Ovine, Open

Reproducibility

Within Run (n=30 )
Blood, Serum, Plasma, Dialysate
Na+: CV  1% (120-160 mmol/ L)
K+: CV  2% (2.5-6 mmol/L )
Cl-: CV  2% (85-130 mmol/ L)
Ca++: SD  0.02 mmol/L (0.8 – 1 .5 mmol/L)
Li+: SD  0.02 mmol/L (0.4 – 1 .3 mmol/L)
(Lithium not typically measured in dialysate samples)
Urine
Na+: CV  5% (100-250 mmol/ L)
K+: CV  5% (10-60 mmol/L )
Cl-: CV  5% (100-250 mmol/ L)
(Calcium and Lithium are not typically measured in urine samples)

Veterinary Options
Feline, Canine, Bovine, Equine, Swine, Ovine, Open

User Input
Keypad, External Keyboard,
External Barcode Scanner (optional)

Output
Graphic backlit display
Graphic 16 column printe r
RS232C serial port

Ambient Conditions
15-32°C, Room temperature (60-90° F)
<85% humidity

Languages
English, Chinese, French, German, Italian ,
Japanese, Portuguese, Russian, Spanish,
Korean, Polish

Power
100-240V ~ 50/60 Hz (self adjustin g)
1.6 A max, 50 Watts max

Size & Weight
13.2 x 12.4 x 12 inches or 335 x 315 x 295 mm
14 lbs or 6 Kgs

User Input
Keypad, External Keyboard,
External Barcode Scanner (optional)

Output
Graphic backlit display
Graphic 16 column printe r
RS232C serial port

Ambient Conditions
15-32°C, Room temperature (60-90° F)
<85% humidity

Languages
English, Chinese, French, German, Italian ,
Japanese, Portuguese, Russian, Spanish,
Korean, Polish

Power
100-240V ~ 50/60 Hz (self adjustin g)
1.6 A max, 50 Watts max

Size & Weight
13.2 x 12.4 x 12 inches or 335 x 315 x 295 mm
14 lbs or 6 Kgs